Experimental Atomic Force Microscopy data of agglomeration and bubble-particle hetero-coagulation

%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%
%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%

Detailed interpretation of the raw AFM data in this repository is found in these research articles: 

https://doi.org/10.1021/acs.langmuir.3c01781	Jan Nicklas, Lisa Ditscherlein and Urs A. Peuker. "Particle–Bubble Interactions: an Investigation of the Three-Phase Contact Line by Atomic Force Microscopy." Langmuir 39.38 (2023): 13630-13640.
https://doi.org/10.1016/j.colsurfa.2024.134660	Jan Nicklas and Urs A. Peuker. "Stability of Bubble-Particle Hetero-Coagulates Accessed by Atomic Force Microscopy to Understand the Froth Flotation of Agglomerating Hydrophobic Fines." Colloids and Surfaces A: Physicochemical and Engineering Aspects (2024)

If you are new to the interpretation of AFM data, the following resource is a good starting point: 
https://doi.org/10.1016/j.surfrep.2005.08.003	Butt, Hans-Jürgen, Brunero Cappella, and Michael Kappl. "Force measurements with the atomic force microscope: Technique, interpretation and applications." Surface science reports 59.1-6 (2005): 1-152.

All measurements were carried out with silanized alumina colloidal probes. The spherical alumina particles (DAW-07, Denka, Japan) were hydrophobized by silanization with the silane 1H,1H,2H,2H-Perfluoro
octyltriethoxysilane (AB104055, C14H19F13O3Si, abcr, Germany). 

The following data sets of base-line corrected raw AFM-data are included in the folder "...\AFM_data":

3 x particle - bubble interaction 
3 x particle-laden bubble - particle interaction
1 x agglomerate - particle interaction

The numbered files "measurement_*.txt" are measurements of the force F as a function of the piezo displacement dX
for the interaction between a colloidal probe particle and an agglomerate or a concentrically aligned bubble or particle-laden bubble.

1. column 	dX_t	µm	piezo displacement on approach (trace)
2. column 	F_t	nN	measured force on approach (trace)
3. column 	dX_r	µm	piezo displacement on retraction (retrace)
4. column 	F_r	nN	measured force on retraction (retrace)

Each dataset includes a metadata file "info_*.txt" with the following information:
- interaction type
- cantilever spring constant k_c
- colloidal probe radius R_CP
- bubble radius R_b (if applicable)
- colloidal probe material
- electroly solution the measurement was taken in
- piezo displacement velocity v
- comment (if applicable)

AFM:
Measurements were taken with an XE100 AFM (Park Systems, Korea) that was equipped with the long-range scan head (25 µm scan range) and a liquid probe hand. 
The measurements were taken in a cylindrical dish (~3 µL volume) filled with electrolyte solution.

Base-line correction:
The dX-F raw data was corrected in Matlab by a linear first order correction. 
The slope for this correction was determined by fitting straight line to the first 100-2000 data points of the measurements. The number of data points 
used in the correction is the same for each "measurement_*.txt" of a dataset. 