Ditscherlein, Ralf2020-05-082026-05-202020-05-082026-05-2020202020-05-08https://opara.zih.tu-dresden.de/handle/123456789/2375https://doi.org/10.25532/OPARA-82To get good image segmentation results in tomographic analysis methods, particles has to be seperated from each other while embedding them into a matrix. Low X-ray absorbing nano-graphite below the voxel resolution of the system is mixed with the analysis particles to form a shell to avoid direct particle-particle contact. After mixing it with epoxy resin, the paste is sucked into a polymeric tube of 2 mm inner diameter. After hardening, the formed cylinder was scanned with an X-ray microscope (ZEISS Xradia Versa 510) with the following parameters: 80 keV, 7 W, 360°, 2001 projections, 1 s exposure time, 2 µm voxel size, binning 2. After scannning, the following filters were applied while reconstruction with ZEISS reconstructor: Beam Hardening Correction 0.05, Smoothing 0,7. Detailed overview of the parameters is summarized as screenshots within the repository. To determine an optimal mixture of the two components, the following volume-fractions were analysed: Nanoparticle:Particle 20vol-%:80vol-% 40vol-%:60vol-% 60vol-%:40vol-%Attribution 4.0 Internationalhttp://creativecommons.org/licenses/by/4.0/4::424Epoxy-Based Particle Sample Preparation with Nano-Graphite as Spacer - RAW Data (Particle Sample Preparation) - Different mixtures (by volume) of graphite-nanoparticles and analysis particles